TEST FUNCTION NAME

Teledyne API - T300/T300M and M300E/EM PN 04906H (DCN5840)APPENDIX A-3: Warnings and Test Functions

Table A-3: T300/T300M and M300E/EM Test Functions, Revision L.8

TEST FUNCTION NAME

MESSAGE TEXT

DESCRIPTION

RANGE

RANGE=50.0 PPM 3

 

D/A range in single or auto-range modes.

 

CO RANGE=50.0 PPM 3, 7

 

RANGE1

RANGE1=50.0 PPM 3

 

D/A #1 range in dual range mode.

 

CO RANGE1=50.0 PPM 3, 7

 

RANGE2

RANGE2=50.0 PPM 3

 

D/A #2 range in dual range mode.

 

CO RANGE2=50.0 PPM 3, 7

 

CO2RANGE

CO2 RANGE=20 % 7

 

CO2 range.

O2RANGE

O2 RANGE=100 % 10

 

O2 range.

STABILITY

STABIL=0.0 PPM 3

 

 

Concentration stability (standard deviation

 

CO STB=0.0 PPM

3, 7, 10

based on setting of STABIL_FREQ and

 

 

 

 

STABIL_SAMPLES).

 

CO2 STB=0.0 % 7

 

 

 

 

 

 

 

 

 

O2 STB=0.0 % 10

 

 

 

 

RESPONSE 2

RSP=0.20(0.00) SEC

 

Instrument response. Length of each

 

 

 

 

 

signal processing loop. Time in

 

 

 

 

 

parenthesis is standard deviation.

COMEAS

CO MEAS=4125.0 MV

Detector measure reading.

COREF

CO REF=3750.0 MV

 

Detector reference reading.

MRRATIO

MR RATIO=1.100

 

 

 

Measure/reference ratio.

AUTOZERO 4, 5

AZERO RATIO=1.234

Measure/reference ratio during auto-

 

 

 

 

 

reference.

SAMPPRESS

PRES=29.9 IN-HG-A

 

Sample pressure.

 

 

 

 

 

 

PURGEPRESS 9

PURGE=7.5 PSIG

 

 

 

Purge pressure

VACUUM 8

VAC=6.8 IN-HG-A

 

 

 

Vacuum pressure.

SAMPFLOW 6

SAMP FL=751 CC/M

 

Sample flow rate.

SAMPTEMP

SAMPLE TEMP=26.8 C

Sample temperature.

BENCHTEMP

BENCH TEMP=48.1 C

Bench temperature.

WHEELTEMP

WHEEL TEMP=68.1 C

Wheel temperature.

O2CELLTEMP 10

O2 CELL TEMP=50.2 C

O2 sensor cell temperature.

BOXTEMP

BOX TEMP=26.8 C

 

 

Internal box temperature.

 

 

 

 

BOXTEMP2 4

BOX TEMP2=29.6 C

 

Internal box temperature #2.

OVENTEMP 11

OVEN TEMP=30.1 C

 

Oven temperature

PHOTOTEMP

PHT DRIVE=2500.0 MV

Photometer temperature.

COSLOPE

SLOPE=1.000

 

 

 

CO slope for current range, computed

 

CO SLOPE=1.000

7

 

 

during zero/span calibration.

 

 

 

 

 

COSLOPE1

SLOPE1=1.000

 

 

 

CO slope for range #1 in dual range

 

CO SLOPE1=1.000

7

 

mode, computed during zero/span

 

 

 

calibration.

 

 

 

 

 

COSLOPE2

SLOPE2=1.000

 

 

 

CO slope for range #2 in dual range

 

CO SLOPE2=1.000

7

 

mode, computed during zero/span

 

 

 

calibration.

 

 

 

 

 

COOFFSET

OFFSET=0.000

 

 

 

CO offset for current range, computed

 

CO OFFSET=0.000

7

 

during zero/span calibration.

 

 

 

 

COOFFSET1

OFFSET1=0.000

 

 

 

CO offset for range #1 in dual range

 

CO OFFSET1=0.000

7

mode, computed during zero/span

 

 

calibration.

 

 

 

 

 

A-23

Page 345
Image 345
Teledyne M300EM operation manual A-23, Test Function Name, Message Text, Description, Stabil_Samples