Teledyne operation manual 7:M300E Family Physical range by Model, Range, M300EM

Models: M300E M300EM

1 398
Download 398 pages 30.02 Kb
Page 99
Image 99
Table 6-7: M300E Family Physical range by Model

Teledyne API – Technical Manual - Model 300E Family CO Analyzers

Basic Operation

6.6.SETUP RNGE: ANALOG OUTPUT REPORTING RANGE CONFIGURATION

6.6.1.PHYSICAL RANGE VERSUS ANALOG OUTPUT REPORTING RANGES

Functionally, the M300E Family of CO Analyzers have one hardware PHYSICAL RANGE that is capable of determining CO concentrations between across a very wide array of values.

Table 6-7: M300E Family Physical range by Model

MODEL

RANGE

M300E

0 – 1000 ppm

M300EM

0 – 5000 ppm

 

 

This architecture improves reliability and accuracy by avoiding the need for extra, switchable, gain-amplification circuitry. Once properly calibrated, the analyzer’s front panel will accurately report concentrations along the entire span of its physical range.

Because many applications use only a small part of the analyzer’s full physical range, this can create data resolution problems for most analog recording devices. For example, in an application where an M300E is being used to measure an expected concentration of typically less than 50 ppm CO, the full scale of expected values is only 4% of the instrument’s full 1000 ppm measurement range. Unmodified, the corresponding output signal would also be recorded across only 2.5% of the range of the recording device.

The M300E/EM Analyzers solve this problem by allowing the user to select a scaled reporting range for the analog outputs that only includes that portion of the physical range relevant to the specific application.

Only this REPORTING RANGE of the analog outputs is scaled, the physical range of the analyzer and the readings displayed on the front panel remain unaltered.

NOTE

Both the iDAS values stored in the CPU’s memory and the concentration values reported on the front

panel are unaffected by the settings chosen for the reporting range(s) of the instrument.

04288D DCN5752

97

Page 99
Image 99
Teledyne operation manual 7:M300E Family Physical range by Model, Range, M300EM