TEST CHANNEL

Advanced Features

Teledyne API – Technical Manual - Model 300E Family CO Analyzers

7.4.6.SELECTING A TEST CHANNEL FUNCTION FOR OUTPUT A4

The test functions available to be reported are listed in Table 7-10:

Table 7-10: Test Channels Functions available on the M300E/EM’s Analog Output

TEST CHANNEL

DESCRIPTION

ZERO

FULL SCALE *

 

 

 

 

NONE

TEST CHANNEL IS TURNED OFF.

 

 

CO MEASURE

The demodulated, peak IR detector output

 

 

during the measure portion of the GFC Wheel

0 mV

5000 mV

 

cycle.

 

 

CO REFERENCE

The demodulated, peak IR detector output

 

 

during the reference portion of the GFC

0 mV

5000 mV

 

Wheel cycle.

 

 

SAMPLE PRESS

The absolute pressure of the Sample gas as

 

 

measured by a pressure sensor located inside

0" Hg

40 "Hg

 

the sample chamber.

 

 

SAMPLE FLOW

Sample mass flow rate as measured by the

0 cm3/m

1000 cm 3/m

 

flow rate sensor in the sample gas stream.

 

 

SAMPLE TEMP

The temperature of the gas inside the sample

0°C

70°C

chamber.

 

 

 

BENCH TEMP

Optical bench temperature.

0°C

70°C

 

 

 

 

WHEEL TEMP

GFC Wheel temperature.

0°C

70°C

O2 CELL TEMP

The current temperature of the O2 sensor

n

70°C

measurement cell.

 

 

 

CHASSIS TEMP

The temperature inside the analyzer chassis.

0°C

70°C

PHT DRIVE

The drive voltage being supplied to the

 

 

thermoelectric coolers of the IR photo-

0 mV

5000 mV

 

detector by the Sync/Demod Board.

 

 

* Maximum test signal value at full scale of test channel output.

Once a function is selected, the instrument not only begins to output a signal on the analog output, but also adds TEST to the list of test functions viewable via the front panel display.

144

04288D DCN5752

Page 146
Image 146
Teledyne M300EM operation manual Test Channel, Description, Zero, Full Scale, None