3.5.2.WARM UP

Getting Started

Teledyne API – Technical Manual - Model 300E Family CO Analyzers

3.5.2.WARM UP

The M300E/EM requires about 60 minutes warm-up time before reliable CO measurements can be taken. During that time, various portions of the instrument’s front panel will behave as shown in Table 3-8.See Figure 3-1for the layout.

Table 3-8: Front Panel Display during System Warm-Up

 

NAME

 

COLOR

 

BEHAVIOR

 

SIGNIFICANCE

 

Concentration

 

 

 

Displays current,

 

 

 

 

N/A

 

compensated CO

 

This is normal operation.

 

Field

 

 

 

 

 

 

 

Concentration

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Mode Field

 

N/A

 

Displays blinking

 

Instrument is in sample mode but is still in the process of

 

 

 

“SAMPLE”

 

warming up. (iDAS holdoff period is active)

 

 

 

 

 

 

 

STATUS LED’s

 

 

 

 

 

 

 

 

 

 

 

 

 

Unit is operating in sample mode; front panel display is being

 

Sample

 

Green

 

On

 

updated.

 

 

 

 

 

 

 

Flashes On/Off when adaptive filter is active

 

Cal

 

Yellow

 

Off

The instrument’s calibration is not enabled.

 

 

 

 

 

 

 

The analyzer is warming up and hence out of specification

 

Fault

Red

 

Blinking

 

for a fault-free reading. Various warning messages will

 

 

 

 

 

 

 

appear.

3.5.3.WARNING MESSAGES

Because internal temperatures and other conditions may be outside the specified limits during the analyzer’s warm-up period, the software will suppress most warning conditions for 30 minutes after power up. If warning messages persist after the 60 minutes warm-up period is over, investigate their cause using the troubleshooting guidelines in Section Error! Reference source not found..

To view and clear warning messages, press:

Table 3-6 lists brief descriptions of the warning messages that may occur during start up.