APPENDIX A-2: Setup Variables For Serial I/O

Setup Variable

Numeric

Default

Value

 

Units

Value

Range

REF_SOURCE_LIMIT

mV

3000 (not

1–5000

 

 

used)

 

 

 

Warnings:

 

 

 

1100–4800,

 

 

 

25–4800 3, 4, 15

 

FACTORY_OPT

BitFlag

512,

0–65535

 

 

768 5

 

 

 

 

 

Description

Reference source warning limits. Set point is not used.

Factory option flags. Add values to combine flags.

1 = enable dilution factor

2 = zero/span valves installed

4 = enable conc. alarms

8 = enable linearity adjustment factor

16 = display units in concentration field

32 = enable software-controlled maintenance mode

64 3, 5 = span valve installed

128 = enable switch-controlled maintenance mode

256 = compute only offset during zero calibration

512 = 220 V A/C power

1024 = non-zero offset calibration (linearity adjustment must also be enabled)

2048 = enable Internet option 7

4096 = use “old” style numeric data entry menus when editing conc. table

8192 = locate high range and zero cal. status outputs on relays