A-16

APPENDIX A-2: Setup Variables For Serial I/O

Teledyne API - T300/T300M and M300E/EM PN 04906H (DCN5840)

Setup Variable

Numeric

Default

Value

 

Units

Value

Range

SAMP_FLOW_SET

cc/m

800,

0–5000

 

 

2000 13

 

 

 

1800 5,19

 

 

 

Warnings:

 

 

 

640–960,

 

 

 

1400–2200

 

 

 

5,19

 

 

 

1500–2500 13

 

SAMP_FLOW_SLOPE

1

0.001–100

 

 

4.5 5,19

 

VAC_SAMP_RATIO

0.53,

0.1–2

 

 

0.61 13

 

 

 

 

 

PURGE_PRESS_SET

PSIG

7.5

0–100

 

 

Warnings:

 

 

 

2.5–12.5

 

SAMP_TEMP_SET

ºC

30

0–100

 

 

Warnings:

 

 

 

10.1–100

 

 

 

 

 

BOX_SET

ºC

30

0–100

 

 

 

 

 

 

Warnings:

 

 

 

5–48

 

BOX2_SET 5,

ºC

30

0–100

OVEN_SET 19,23

 

46 19,23

 

 

 

Warnings:

 

 

 

25–35

 

 

 

41–51 19,23

 

BOX2_CYCLE 5,

Seconds

10

0.5–30

OVEN_CYCLE 19,23

 

 

 

BOX2_PROP 5,

1/ºC

1

0–100

OVEN_PROP 19,23

 

0.5 19,23

 

 

 

 

 

BOX2_INTEG 5,

0.1

0–100

OVEN_INTEG 19,23

 

0.02 19,23

 

BOX2_DERIV 5,

0

0–100

OVEN_DERIV 19,23

 

 

 

BENCH_CYCLE

Seconds

2

0.5–30

 

 

15 19,23

 

BENCH_PROP

1/ºC

5

0–100

 

 

1.5 19,23

 

 

 

 

 

BENCH_INTEG

0.5

0–100

 

 

1.5 19,23

 

Description

Sample flow warning limits. Set point is not used.

Slope term to correct sample flow rate.

Maximum vacuum pressure / sample pressure ratio for valid sample flow calculation.

Purge pressure warning limits. Set point is not used.

Sample temperature warning limits. Set point is not used.

Internal box temperature warning limits. Set point is not used.

Internal box temperature #2 / oven set point and warning limits.

Internal box temperature #2/oven control cycle period.

Internal box temperature #2/oven PID proportional coefficient. Proportional band is the reciprocal of this setting.

Internal box temperature #2/oven PID integral coefficient.

Internal box temperature #2/oven PID derivative coefficient.

Optical bench temperature control cycle period.

100V optical bench temperature PID proportional coefficient. Proportional band is the reciprocal of this setting.

100V optical bench temperature PID integral coefficient.

A-16

04188D DCN5752

Page 338
Image 338
Teledyne M300EM operation manual Setup Variable, Numeric, Default, Value, Units, Range, Description, A-16