Image 90
6.3.WARNING MESSAGES

Basic Operation

Teledyne API – Technical Manual - Model 300E Family CO Analyzers

6.3.WARNING MESSAGES

The most common instrument failures will be reported as a warning on the analyzer’s front panel and through the COMM ports. Section 13.1.1 explains how to use these messages to troubleshoot problems. Section 6.3 shows how to view and clear warning messages.

Table 6-3: List of Warning Messages

MESSAGE

MEANING

 

 

ANALOG CAL WARNING

The instrument’s A/D circuitry or one of its analog outputs is not calibrated.

BENCH TEMP WARNING

The temperature of the optical bench is outside the specified limits.

BOX TEMP WARNING

The temperature inside the chassis is outside the specified limits.

CANNOT DYN SPAN2

Remote span calibration failed while the dynamic span feature was set to turned on.

CANNOT DYN ZERO3

Remote zero calibration failed while the dynamic zero feature was set to turned on.

CONC ALRM1 WARNING1

Concentration alarm 1 is enabled and the measured CO level is ≥ the set point.

CONC ALRM2 WARNING1

Concentration alarm 2 is enabled and the measured CO level is ≥ the set point.

CONFIG INITIALIZED

Configuration storage was reset to factory configuration or erased.

DATA INITIALIZED

iDAS data storage was erased.

O2 CELL TEMP WARN2

O2 sensor cell temperature outside of warning limits.

PHOTO TEMP WARNING

The temperature of the IR photo detector is outside the specified limits.

REAR BOARD NOT DET

The CPU is unable to communicate with the motherboard.

RELAY BOARD WARN

The firmware is unable to communicate with the relay board.

SAMPLE FLOW WARN

The flow rate of the sample gas is outside the specified limits.

SAMPLE PRESS WARN

Sample gas pressure outside of operational parameters.

SAMPLE TEMP WARN

The temperature of the sample gas is outside the specified limits.

SOURCE WARNING

The IR source may be faulty.

SYSTEM RESET1

The computer was rebooted.

WHEEL TEMP WARNING

The Gas Filter Correlation Wheel temperature is outside the specified limits.

1Alarm warnings only present when 0ptional alarm package is activated.

2Only enabled when the optional O2 Sensor is installed.