3.6.2.3.Set CO Span Gas Concentration

Getting Started

Teledyne API – Technical Manual - Model 300E Family CO Analyzers

3.6.2.3.Set CO Span Gas Concentration

Set the expected CO pan gas concentration. This should be 80-90% of range of concentration range for which the analyzer’s analog output range is set.

54