Teledyne M300EM operation manual Set CO Span Gas Concentration

Models: M300E M300EM

1 398
Download 398 pages 30.02 Kb
Page 56
Image 56
3.6.2.3.Set CO Span Gas Concentration

Getting Started

Teledyne API – Technical Manual - Model 300E Family CO Analyzers

3.6.2.3.Set CO Span Gas Concentration

Set the expected CO pan gas concentration. This should be 80-90% of range of concentration range for which the analyzer’s analog output range is set.

54

04288D DCN5752

Page 56
Image 56
Teledyne M300EM operation manual Set CO Span Gas Concentration