Teledyne M300EM operation manual Using the dilution ratio option is a 4-stepprocess

Models: M300E M300EM

1 398
Download 398 pages 30.02 Kb
Page 107
Image 107
Using the dilution ratio option is a 4-step process:

Teledyne API – Technical Manual - Model 300E Family CO Analyzers

Basic Operation

6.6.5.SETUP RNGE DIL: USING THE OPTIONAL DILUTION RATIO FEATURE

This feature is a optional software utility that allows the user to compensate for any dilution of the sample gas that may occur before it enters the sample inlet. Typically this occurs in continuous emission monitoring (CEM) applications where the sampling method used to remove the gas from the stack dilutes it.

Using the dilution ratio option is a 4-step process:

1.Select the appropriate units of measure (see Section 6.6.4).

2.Select the reporting range mode and set the reporting range upper limit (see Section 6.6.3). Make sure that:

The upper span limit entered for the reporting range is the maximum expected concentration of the UNDILUTED gas.

3.Set the dilution factor as a gain (e.g., a value of 20 means 20 parts diluent and 1 part of sample gas):

4.Calibrate the analyzer.

Make sure that the calibration span gas is either supplied through the same dilution system as the sample gas or has an appropriately lower actual concentration.

EXAMPLE: If the reporting range limit is set for 100 ppm and the dilution ratio of the sample gas is 20 gain, either:

a span gas with the concentration of 100 ppm can be used if the span gas passes through the same dilution steps as the sample gas, or;

a 5 ppm span gas must be used if the span gas IS NOT routed through the dilution system.

04288D DCN5752

105

Page 107
Image 107
Teledyne M300EM operation manual Using the dilution ratio option is a 4-stepprocess