Teledyne API – Technical Manual - Model 300E Family CO Analyzers

Calibration Procedures

 

 

CONTINUED FROM PREVIOUS PAGE

 

 

 

 

DELTA TIME

 

 

 

 

 

 

 

 

SETUP X.X

DURATION:15.0 MINUTES

 

 

 

<SET SET>

EDIT

 

EXIT

 

 

 

 

 

 

 

 

 

 

 

 

Toggle keys to set

SETUP X.X

DURATION 15.0MINUTES

 

 

duration for each iteration

 

 

 

 

 

 

 

of the sequence:

3

0 .0

ENTR

EXIT

Set in Decimal minutes

from 0.1 – 60.0.

 

 

 

 

 

 

 

 

 

 

 

SETUP X.X

DURATION:30.0 MINUTES

 

 

 

<SET SET>

EDIT

 

EXIT

 

 

 

 

 

 

Toggle key

Between Off and

ON.

Display show:

SEQ 2) ZERO–SPAN, 2:00:30

Sequence

Delta Time

MODE

Delta Days

 

SETUP X.X

CALIBRATE: OFF

 

 

 

<SET SET> EDIT

EXIT

 

 

 

 

 

 

 

 

 

 

 

SETUP X.X

CALIBRATE: OFF

 

 

 

ON

ENTR

EXIT

 

 

 

 

 

 

 

 

 

 

 

SETUP X.X

CALIBRATE: ON

 

 

 

<SET SET>

EDIT

EXIT

 

 

 

 

 

 

 

 

 

 

 

SETUP X.X

SEQ 2) ZERO–SPAN, 2:00:30

 

 

 

PREV NEXT MODE SET

EXIT

 

 

 

 

 

 

EXIT returns to the SETUP Menu.

9.4.1.1.AutoCal with Auto or Dual Reporting Ranges Modes Selected

If the M300E/EM Analyzer is set for either the Dual or Auto reporting range modes, the following three steps will appear at the beginning of the AutoCal setup routine:

SETUP X.X

RANGE TO CAL: LOW

 

<SET

EDIT

EXIT

 

 

 

 

 

 

SETUP X.X

RANGE TO CAL: LOW

 

LOW HIGH

ENTR

SETUP

 

 

 

 

 

 

SETUP X.X

RANGE TO CAL: HIGH

 

<SET

EDIT

EXIT

 

 

 

 

 

 

SETUP X.X

SEQ 2) ZERO–SPAN, 2:00:30

 

PREV NEXT MODE SET

EXIT

 

 

 

EXIT returns to the

PRIMARY SETUP

Menu.

NOTE

In order to automatically calibrate both the HIGH and LOW ranges, you must set up a separate sequence

for each.

04288D DCN5752

203

Page 205
Image 205
Teledyne M300EM operation manual for each