Teledyne M300EM operation manual Epa Equivalency Designation

Models: M300E M300EM

1 398
Download 398 pages 30.02 Kb
Page 30
Image 30

Specifications and Approvals

Teledyne API – Technical Manual - Model 300E Family CO Analyzers

2.2.EPA EQUIVALENCY DESIGNATION

Teledyne API’s M300E Carbon Monoxide Analyzer is designated as Reference Method Number EQOA-0992-

087 as defined in 40 CFR Part 53, when operated under the following conditions:

Range: Any range from 10 ppm to 50 ppm.

Ambient temperature range of 10 to 40°C.

Line voltage range of 90 – 127 and 200 – 230 VAC, 50/60 Hz.

Sample filter: Equipped with PTFE filter element in the internal filter assembly.

Sample flow of 800 ± 80 cm3/min at sea level.

Internal sample pump.

Software settings:

Dilution factor

1.0

AutoCal

ON or OFF

Dynamic Zero

ON or OFF

Dynamic Span

OFF

Dual range

ON or OFF

Auto range

ON or OFF

 

 

Temp/Pres compensation

ON

 

 

Under the designation, the analyzer may be operated with or without the following options:

Rack mount with slides.

Rack mount without slides, ears only.

Zero/span valve options.

Option 50A – Sample/Cal valves, or;

Option 50B – Sample/Cal valves with span shutoff & flow control.

Internal zero/span (IZS) option with either:

Option 51A – Sample/Cal valves, or;

Option 51C – Sample/Cal valves with span shutoff & flow control.

Status outputs.

Control inputs.

RS-232 output.

Ethernet output.

4-20mA, isolated output.

28

04288D DCN5752

Page 30
Image 30
Teledyne M300EM operation manual Epa Equivalency Designation