9.7.CALIBRATION OF OPTIONAL SENSORS

Teledyne API – Technical Manual - Model 300E Family CO Analyzers

Calibration Procedures

9.7.CALIBRATION OF OPTIONAL SENSORS

9.7.1.O2 SENSOR CALIBRATION PROCEDURE

9.7.1.1.O2 Calibration Setup

The pneumatic connections for calibrating are as follows:

Figure 9-7: O2 Sensor Calibration Set Up

O2 SENSOR ZERO GAS: Teledyne API recommends using pure N2 when calibration the zero point of your O2 sensor option.

O2 SENSOR SPAN GAS: Teledyne API recommends using 20.8% O2 in N2 when calibration the span point of your O2 sensor option (See Table 3-7).

04288D DCN5752

209

Page 211
Image 211
Teledyne M300E Calibration Of Optional Sensors, 9.7.1.O2 SENSOR CALIBRATION PROCEDURE, 9.7.1.1.O2 Calibration Setup