Teledyne M300EM ZERO/SPAN/SHUTOFF VALVE OPTION 50B, Internal Pneumatics OPT 50B, Figure

Models: M300E M300EM

1 398
Download 398 pages 30.02 Kb
Page 69
Image 69
5.6.3.ZERO/SPAN/SHUTOFF VALVE (OPTION 50B)

Teledyne API – Technical Manual - Model 300E Family CO Analyzers

Optional Hardware and Software

5.6.3.ZERO/SPAN/SHUTOFF VALVE (OPTION 50B)

This option requires that both zero air and span gas be supplied from external sources.

Span gas will be supplied from a pressurized bottle of calibrated CO gas.

A critical flow control orifice, internal to the instrument ensures that the proper flow rate is maintained.

An internal vent line ensures that the gas pressure of the span gas is reduced to ambient atmospheric pressure.

A SHUTOFF valve preserves the span gas source when it is not in use.

Zero gas is supplied by either an external scrubber or a zero air generator such as the Teledyne API’s M701.

5.6.3.1.Internal Pneumatics (OPT 50B)

Figure 5-5:

Internal Pneumatic Flow OPT 50B – Zero/Span/Shutoff Valves

 

Table 5-2: Zero/Span Valve Operating States for Option 50B

 

 

 

 

 

 

 

MODE

 

VALVE

CONDITION

 

 

SAMPLE

Sample/Cal

Open to SAMPLE inlet

 

 

(Normal

 

Zero/Span

Open to IZS inlet

 

 

State)

 

 

 

 

 

 

Shutoff Valve

Closed

 

 

 

 

 

 

 

 

Sample/Cal

Open to ZERO/SPAN valve

 

 

ZERO CAL

Zero/Span

Open to IZS inlet

 

 

 

 

 

 

 

 

 

 

Shutoff Valve

Closed

 

 

 

 

 

 

 

 

 

 

Sample/Cal

Open to ZERO/SPAN valve

 

 

SPAN CAL

Zero/Span

Open to SHUTOFF valve

 

 

 

 

Shutoff Valve

Open to PRESSURE SPAN Inlet

 

04288D DCN5752

67

Page 69
Image 69
Teledyne M300EM operation manual ZERO/SPAN/SHUTOFF VALVE OPTION 50B, Internal Pneumatics OPT 50B, Figure