Table 7-7: Analog Output Voltage Range Min/Max

Teledyne API – Technical Manual - Model 300E Family CO Analyzers

Advanced Features

7.4.2.ANALOG OUTPUT VOLTAGE / CURRENT RANGE SELECTION

In its standard configuration, each of the analog outputs is set to output a 0–5 VDC signals. Several other output ranges are available. Each range has is usable from -5% to + 5% of the rated span.

Table 7-7: Analog Output Voltage Range Min/Max

RANGE NAME

RANGE SPAN

 

MINIMUM OUTPUT

MAXIMUM OUTPUT

 

 

 

 

 

0.1V

0-100 mVDC

 

-5 mVDC

105 mVDC

 

 

 

 

 

1V

0-1 VDC

 

-0.05 VDC

1.05 VDC

 

 

 

 

 

5V

0-5 VDC

 

-0.25 VDC

5.25 VDC

 

 

 

 

 

10V

0-10 VDC

 

-0.5 VDC

10.5 VDC

 

 

 

 

 

The default offset for all VDC ranges is 0-5 VDC.

 

 

 

 

 

 

 

CURR

0-20 mA

 

0 mA

20 mA

 

 

 

 

 

While these are the physical limits of the current loop modules, typical applications use 2-20 or 4-20 mA for the lower and upper limits. Please specify desired range when ordering this option.

The default offset for all current ranges is 0 mA. Current outputs are available only on A1-A3.

04288D DCN5752

131

Page 133
Image 133
Teledyne M300EM 7:Analog Output Voltage Range Min/Max, Range Name, Range Span, Minimum Output, Maximum Output