Teledyne API – Technical Manual - Model 300E Family CO Analyzers

Getting Started

3.6.2.4.Zero/Span Calibration

To perform the zero/span calibration procedure, press:

SAMPLE

RANGE=0.0 PPm

CO= XX.XX

< TST TST >

CAL

SETUP

 

 

 

 

 

 

 

 

 

 

 

 

Toggle TST> button until ...

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SAMPLE

STABIL= XXXX PPM

CO=XX.XX

< TST TST >

CAL

SETUP

 

 

 

 

 

 

 

 

 

 

Set the Display to show the STABIL test function. This function calculates the stability of the CO measurement.

Allow zero gas to enter the sample port

at the rear of the analyzer.

Wait until STABIL falls below 0.5 ppm.

This may take several

minutes.

SAMPLE

STABIL= XXXX PPM

CO=XX.XX

< TST TST >

CAL

SETUP

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

M-P CAL

STABIL= XXXX PPM

CO=XX.XX

<TST TST>

ZERO CONC

EXIT

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

M-P CAL

STABIL= XXXX PPM

CO=XX.XX

<TST TST> ENTR

CONC

EXIT

 

 

 

 

 

 

 

 

 

 

 

 

Press ENTR to changes the OFFSET & SLOPE values for the CO measurements.

STABIL= XXXX PPM Press EXIT to leave the calibration unchanged and

return to the previous

menu.

Allow span gas to enter the sample port

at the rear of the analyzer.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Wait until STABIL

 

 

 

 

 

 

 

 

 

 

 

 

 

 

falls below 0.5 PPM.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

This may take several

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

minutes.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SAMPLE

STABIL= XXXX PPM

CO=XX.XX

 

 

 

 

 

 

The SPAN key now appears

 

< TST TST >

CAL

 

SETUP

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

during the transition from

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

zero to span.

 

 

 

 

 

 

 

 

 

 

 

 

 

Press ENTR to changes

You may see both keys.

 

M-P CAL

STABIL= XXXX PPM

CO=XX.XX

 

 

 

 

 

 

 

 

 

 

 

the OFFSET & SLOPE

If either the ZERO or SPAN

 

<TST TST>

ZERO SPAN CONC

EXIT

 

 

 

 

 

values for the CO

 

 

 

 

 

 

 

 

measurements.

buttons fail to appear see the

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Troubleshooting section for

 

 

 

 

 

 

 

 

 

 

 

 

 

Press EXIT to leave the

 

 

 

 

 

 

 

 

 

 

 

 

 

tips.

 

 

 

 

 

 

 

 

 

 

 

 

calibration unchanged and

 

M-P CAL

STABIL= XXXX PPM

CO=XX.XX

 

 

 

 

 

 

 

 

 

 

 

return to the previous

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

<TST TST>

ENTR

CONC

EXIT

 

 

 

 

 

menu.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

M-P CAL

STABIL= XXXX PPM

CO=XX.XX

 

 

 

 

 

EXIT at this point

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

<TST TST>

ENTR

CONC

EXIT

 

 

 

 

 

returns to the

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SAMPLE menu.

 

 

 

 

 

 

 

 

 

 

 

 

 

 