and alters the instrument’s Calibration.

Calibration Procedures

Teledyne API – Technical Manual - Model 300E Family CO Analyzers

9.2.MANUAL CALIBRATION CHECKS AND CALIBRATION OF THE M300E/EM ANALYZER IN ITS BASE CONFIGURATION

ZERO/SPAN CALIBRATION CHECKS VS. ZERO/SPAN CALIBRATION

Pressing the ENTR key during the following procedure resets the stored values for OFFSET and SLOPE

and alters the instrument’s Calibration.

This should ONLY BE DONE during an actual calibration of the M300E/EM.

NEVER press the ENTR key if you are only checking calibration.

9.2.1.SETUP FOR BASIC CALIBRATION CHECKS AND CALIBRATION

STEP ONE: Connect the Sources of Zero Air and Span Gas as shown below.

Figure 9-1: Pneumatic Connections – Basic Configuration – Using Bottled Span Gas

188

04288D DCN5752

Page 190
Image 190