9.2.3.PERFORMING A BASIC MANUAL CALIBRATION

Teledyne API – Technical Manual - Model 300E Family CO Analyzers

Calibration Procedures

9.2.3.PERFORMING A BASIC MANUAL CALIBRATION

The following section describes the basic method for manually calibrating the M300E/EM.

If the analyzer’s reporting range is set for the AUTO range mode, a step will appear for selecting which range is to be calibrated (LOW or HIGH). Each of these two ranges MUST be calibrated separately.

9.2.3.1.Setting the Expected Span Gas Concentration

NOTE

When setting expected concentration values, consider impurities in your span gas.

The expected CO span gas concentration should be 80% of the reporting range of the instrument (see Section 6.6.1).

The default factory setting is 40 ppm. To set the span gas concentration, press:

 

 

 

 

SAMPLE

RANGE=50.0PPM

CO=XX.XX

 

 

 

 

 

 

 

 

< TST

CAL

MSG

SETUP

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Only appears if either

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SAMPLE

 

 

 

 

 

 

 

 

 

 

 

 

 

the O2 or CO2

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

O2

 

 

 

 

 

 

 

 

EXIT

 

 

 

Sensors are installed.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SAMPLE

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Only appears if the

or

 

 

HIGH

 

 

 

 

 

 

 

 

EXIT

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

range modes are selected.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Use these keys to choose the

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

appropriate range.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

M-P CAL

 

RANGE=50.0PPM

CO=XX.XX

 

 

 

 

 

Repeat entire procedure for each

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

range.

 

 

 

<TST TST> ZERO SPAN

 

 

 

 

EXIT

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

The CO span concentration value is

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

ignores the new

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

setting and returns to

automatically default to

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

the previous display.

4

.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

If this is not the the concentration of

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

the span gas being used, toggle

 

 

 

 

 

 

 

 

 

 

 

 

 

 

accepts the new

these buttons to set the correct

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

setting and returns to

concentration of the CO

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

the

 

calibration gas.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

NOTE

For this Initial Calibration it is important to independently verify the PRECISE CO Concentration Value of

the SPAN gas.

If the source of the Span Gas is from a Calibrated Bottle, use the exact concentration value printed on

the bottle.

04288D DCN5752

191

Page 193
Image 193
Teledyne M300EM operation manual Performing A Basic Manual Calibration, the SPAN gas, the bottle