Table 6-2:

 

Teledyne API – Technical Manual - Model 300E Family CO Analyzers

Basic Operation

 

 

 

 

Table 6-2:

Test Functions Defined

 

 

 

 

 

 

 

 

 

 

PARAMETER

DISPLAY TITLE

UNITS

MEANING

 

 

 

 

 

PPB3, PPM

Standard deviation of CO concentration readings. Data points are

 

 

Stability

STABIL

recorded every ten seconds using the last 25 data points. This

 

 

UGM3, MGM

function can be reset to show O2 or CO2 stability in instruments with

 

 

 

 

 

those sensor options installed.

 

 

 

 

RANGE

PPB, PPM,

The full scale limit at which the reporting range of the analyzer is

 

 

Range

1

currently set. THIS IS NOT the Physical Range of the instrument.

 

 

RANGE11

UGM, MGM

 

 

 

RANGE2

 

See Section 6.6.1 for more information.

 

 

 

O2 Range 1

O2 RANGE

%

The range setting for the optional O2 Sensor.

 

 

 

CO2 Range2

CO2 RANGE

%

The range setting for the optional CO2 Sensor.

 

 

 

CO Measure

CO MEAS

MV

The demodulated, peak IR detector output during the measure

 

 

portion of the GFC Wheel cycle.

 

 

 

 

 

 

 

 

 

CO Reference

CO REF

MV

The demodulated, peak IR detector output during the reference

 

 

portion of the GFC Wheel cycle.

 

 

 

 

 

 

 

 

 

Measurement /

MR Ratio

 

The result of CO MEAS divided by CO REF. This ratio is the

 

 

-

primary value used to compute CO concentration. The value

 

 

Reference Ratio

 

 

 

 

displayed is not linearized.

 

 

 

 

 

 

 

 

 

Sample Pressure

PRES

In-Hg-A

The absolute pressure of the Sample gas as measured by a

 

 

pressure sensor located inside the sample chamber.

 

 

 

 

 

 

 

Sample Flow

SAMPLE FL

cm3/min

Sample mass flow rate as measured by the flow rate sensor in the

 

 

 

 

 

sample gas stream.

 

 

 

Sample

SAMP TEMP

°C

The temperature of the gas inside the sample chamber.

 

 

Temperature

 

 

 

 

 

 

 

 

Bench

BENCH TEMP

°C

Optical bench temperature.

 

 

 

Temperature

 

 

 

 

 

 

 

 

 

Wheel

WHEEL TEMP

°C

GFC Wheel temperature.

 

 

 

Temperature

 

 

 

 

 

 

 

 

 

Box Temperature

BOX TEMP

°C

The temperature inside the analyzer chassis.

 

 

 

O2 Cell

3

 

 

 

 

 

Temperature3

O2 CELL TEMP

°C

The current temperature of the O2 sensor measurement cell.

 

 

Photo-detector

PHT DRIVE

mV

The drive voltage being supplied to the thermoelectric coolers of the

 

 

Temp. Control

 

 

IR photo-detector by the sync/demod Board.

 

 

 

Voltage

 

 

 

 

 

 

 

 

 

 

 

Slope

SLOPE

-

The sensitivity of the instrument as calculated during the last

 

 

calibration activity.

 

 

 

 

 

 

 

 

 

Offset

OFFSET

-

The overall offset of the instrument as calculated during the last

 

 

calibration activity.

 

 

 

 

 

 

 

 

 

O2 Sensor

O2 SLOPE

-

O2 slope, computed during zero/span calibration.

 

 

Slope 1

 

O2 Sensor Offset 1

O2 OFFSET

-

O2 offset, computed during zero/span calibration.

 

 

CO2 Sensor

CO2 SLOPE

-

CO2 slope, computed during zero/span calibration.

 

 

Slope2

 

 

CO2 Sensor

CO2 OFFSET

-

CO2 offset, computed during zero/span calibration.

 

 

Offset 2

 

 

Current Time

TIME

-

The current time. This is used to create a time stamp on iDAS

 

 

readings, and by the AUTOCAL feature to trigger calibration events.

 

 

 

 

 

 

1Only appears when the optional O2 sensor is installed.

2Only appears when the optional CO2 sensor is installed.

3Only available on the M300E.

04288D DCN5752

87

Page 89
Image 89
Teledyne M300EM operation manual Table, Test Functions Defined