8.6.5.SETTING THE HESSEN PROTOCOL RESPONSE MODE

Remote Operation

Teledyne API – Technical Manual - Model 300E Family CO Analyzers

8.6.5.SETTING THE HESSEN PROTOCOL RESPONSE MODE

The Teledyne API’s implementation of Hessen Protocol allows the user to choose one of several different modes of response for the analyzer.

 

Table 8-7: Teledyne API’s Hessen Protocol Response Modes

 

 

MODE ID

MODE DESCRIPTION

CMD

This is the Default Setting. Reponses from the instrument are encoded as the traditional

command format. Style and format of responses depend on exact coding of the initiating

 

command.

BCC

Responses from the instrument are always delimited with <STX> (at the beginning of the

response, <ETX> (at the end of the response followed by a 2 digit Block Check Code

 

(checksum), regardless of the command encoding.

TEXT

Responses from the instrument are always delimited with <CR> at the beginning and the

end of the string, regardless of the command encoding.

 

To select a Hessen response mode, press:

178

04288D DCN5752

Page 180
Image 180
Teledyne M300EM operation manual Setting The Hessen Protocol Response Mode, Mode Id, Mode Description, Text