APPENDIX A-2: Setup Variables For Serial I/O

Teledyne API - T300/T300M and M300E/EM PN 04906H (DCN5840)

Setup Variable

Numeric

Default

Value

 

Units

Value

Range

USER_UNITS

PPM 0

PPB,

 

 

 

PPM,

 

 

 

UGM,

 

 

 

MGM

 

 

 

% 4, 5, 9, 18

 

 

 

PPM 3, 8

 

 

 

MGM 3, 8

NEG_CONC_SUPPRESS

OFF,

OFF, ON

 

 

ON 17

 

 

 

 

 

DIL_FACTOR

1

0.1–1000

 

 

 

 

DARK_CAL_DURATION

Seconds

180,

10–600

 

 

60 4

 

 

 

 

 

DARK_MEAS_MV

mV

0

-1000–1000

DARK_REF_MV

mV

0

-1000–1000

CO2_COMP_ENABLE

OFF

ON, OFF

 

 

 

 

CO2_COMP_CONC

%

0

0–20

 

 

 

 

SOURCE_DRIFT_ENAB 21

OFF

ON, OFF

 

 

 

 

SOURCE_DRIFT 21

PPB/Day

0

-500–500

 

 

 

 

CO_CONST1

8000,

100–50000

 

 

50015,20,22,23

 

 

 

78.8 9,12

 

 

 

3020 18

 

 

 

500 4,9,12

 

 

 

39600 8

 

 

 

40000 3

 

CO_CONST2

0.2110

0–10

 

 

0.356 20,22,23

 

 

 

0.367 15

 

 

 

1.458 9,12

 

 

 

1.4625 18

 

 

 

1.448 4

 

 

 

0.192 8

 

 

 

0.187 3

 

 

 

0.1196 24

 

ET_MEAS_GAIN

1

0.0001–9.9999

 

 

 

 

ET_REF_GAIN

1

0.0001–9.9999

 

 

 

 

Description

Concentration units for user interface.

ON pegs negative concentrations at zero; OFF permits negative concentrations

Dilution factor. Used only if is dilution enabled with FACTORY_OPT variable.

Duration of dark cal. First two- thirds is stabilization period; final third is measure period.

Dark offset for measure reading. Dark offset for reference reading.

ON enables CO2 compensation; OFF disables it.

CO2 concentration to compensate for.

ON enables source drift compensation; OFF disables it.

Source drift compensation rate of change.

CO calculation constant.

CO calculation constant.

Electrical test gain factor for measure reading.

Electrical test gain factor for reference reading.