Teledyne M300EM operation manual Fault Diagnosis With Test Functions

Models: M300E M300EM

1 398
Download 398 pages 30.02 Kb
Page 272
Image 272
13.1.2.FAULT DIAGNOSIS WITH TEST FUNCTIONS

Troubleshooting & Repair

Teledyne API – Technical Manual - Model 300E Family CO Analyzers

Table 13-1: Warning Messages – Indicated Failures (cont.)

WARNING

FAULT CONDITION

POSSIBLE CAUSES

MESSAGE

 

 

 

 

Ambient temperature outside of specified range

SAMPLE TEMP

Sample temperature is < 10oC or >

Failed bench heater

Failed bench temperature sensor

WARN

100oC.

Relay controlling the bench heater

 

 

Failed relay board

 

 

I2C bus

 

 

GFC Wheel stopped

 

Occurs when CO Ref is <1250

Failed sync/demod board

 

mVDC or >4950 mVDC.

If status LED’s on the sync/demod board ARE flashing the cause is

SOURCE WARNING

 

most likely a failed:

 

IR source

 

Either of these conditions will result

 

Relay board

 

in an invalid M/R ratio.

I2C bus

 

 

IR photo-detector

 

 

This message occurs at power on. If you have not cycled the power

SYSTEM RESET

The computer has rebooted.

on your instrument:

Failed +5 VDC power,

 

 

 

Fatal error caused software to restart

 

 

Loose connector/wiring

 

 

Blocked cooling vents below GFC Assembly. Make sure that

 

 

adequate clear space beneath the analyzer.

WHEEL TEMP

The filter wheel temperature is

Analyzer’s top cover removed

Wheel heater

WARNING

controlled at 68 ± 2 °C

Wheel temperature sensor

 

 

Relay controlling the wheel heater

 

 

Entire relay board

 

 

I2C bus

13.1.2.FAULT DIAGNOSIS WITH TEST FUNCTIONS

Besides being useful as predictive diagnostic tools, the test functions viewable from the front panel can be used to isolate and identify many operational problems when combined with a thorough understanding of the analyzer’s Theory of Operation (see Section Error! Reference source not found.).

The acceptable ranges for these test functions are listed in the “Nominal Range” column of the analyzer Final Test and Validation Data Sheet (M300E, P/N 04307 and M300EM, P/N 04311) shipped with the instrument. Values outside these acceptable ranges indicate a failure of one or more of the analyzer’s subsystems. Functions whose values are still within the acceptable range but have significantly changed from the measurement recorded on the factory data sheet may also indicate a failure.

NOTE:

A worksheet has been provided in Appendix C to assist in recording the value of these test functions.

This worksheet also includes expected values for the various test functions.

270

04288D DCN5752

Page 272
Image 272
Teledyne M300EM operation manual Fault Diagnosis With Test Functions