Teledyne M300EM Setup Variable, Numeric Units, Default Value, Value Range, Description

Models: M300E M300EM

1 398
Download 398 pages 30.02 Kb
Page 330
Image 330
APPENDIX A-2: Setup Variables For Serial I/O

APPENDIX A-2: Setup Variables For Serial I/O

Teledyne API - T300/T300M and M300E/EM PN 04906H (DCN5840)

APPENDIX A-2: Setup Variables For Serial I/O

Table A-1: T300/T300M and M300E/EM Setup Variables, Revision L.8

Setup Variable

Numeric

Units

Default

Value

Value

Range

Description

Low Access Level Setup Variables (818 password)

DAS_HOLD_OFF

Minutes

15

0.5–20

 

Duration of DAS hold off period.

CONC_PRECISION

3

AUTO,

 

Number of digits to display to the

 

 

 

0,

 

 

right of the decimal point for

 

 

 

 

 

concentrations on the display.

 

 

 

1,

 

 

 

 

 

 

 

 

 

 

 

2,

 

 

 

 

 

 

3,

 

 

 

 

 

 

4

 

 

 

REM_CAL_DURATION 17

Minutes

20

1–120

 

Duration of automatic calibration

 

 

 

 

 

 

initiated from TAI protocol.

STABIL_GAS

CO 0

CO,

 

 

Selects gas for stability

 

 

 

CO2

10

,

measurement.

 

 

 

 

 

 

 

 

O2 14

 

 

DYN_ZERO

OFF

ON, OFF

ON enables remote dynamic

 

 

 

 

 

 

zero calibration; OFF disables it.

DYN_SPAN

OFF

ON, OFF

ON enables remote dynamic

 

 

 

 

 

 

span calibration; OFF disables it.

CLOCK_ADJ

Sec./Day

0

-60–60

Time-of-day clock speed

 

 

 

 

 

 

adjustment.

Medium Access Level Setup Variables (929 password)

LANGUAGE_SELECT

ENGL 0

ENGL,

 

 

 

SECD,

 

 

 

EXTN

MAINT_TIMEOUT

Hours

2

0.1–100

 

 

 

 

CONV_TIME

33 MS 0

33 MS,

 

 

 

66 MS,

 

 

 

133 MS,

 

 

 

266 MS,

 

 

 

533 MS,

 

 

 

1 SEC,

 

 

 

2 SEC

 

 

 

 

CO_DWELL

Seconds

0.2

0.1–30

 

 

 

 

CO_SAMPLE

Samples

1

1–30

 

 

 

 

PRE_FILT_SIZE 5, 19

Samples

16

1–50

FILT_SIZE

Samples

750,

1–1000

 

 

720 9, 12

 

 

 

200 3, 8

 

 

 

1000 19, 23

 

Selects the language to use for the user interface.

Time until automatically switching out of software- controlled maintenance mode.

Conversion time for measure/reference detector channel.

Dwell time before taking measure or reference sample.

Number of samples to take in measure or reference mode.

Moving average pre-filter size. Moving average filter size.

A-8

04188D DCN5752

Page 330
Image 330
Teledyne M300EM operation manual Setup Variable, Numeric Units, Default Value, Value Range, Description