8.6.7.SETTING HESSEN PROTOCOL STATUS FLAGS

Remote Operation

Teledyne API – Technical Manual - Model 300E Family CO Analyzers

8.6.7.SETTING HESSEN PROTOCOL STATUS FLAGS

Teledyne API’s implementation of Hessen protocols includes a set of status bits that the instrument includes in responses to inform the host computer of its condition. Each bit can be assigned to one operational and warning message flag. The default settings for these bit/flags are:

Table 8-8: Default Hessen Status Flag Assignments

STATUS FLAG NAME

DEFAULT BIT

ASSIGNMENT

 

WARNING FLAGS

 

SAMPLE FLOW WARNING

0001

BENCH TEMP WARNING

0002

SOURCE WARNING

0004

BOX TEMP WARNING

0008

WHEEL TEMP WARNING

0010

SAMPLE TEMP WARN

0020

SAMPLE PRESS WARN

0040

INVALID CONC

0080

(The Instrument’s Front Panel Display Will Show The

Concentration As “Warnings”)

 

OPERATIONAL FLAGS1

 

Instrument OFF

0100

In MANUAL Calibration Mode

0200

In ZERO Calibration Mode4

0400

In O2 Calibration Mode (if O2 sensor installed )2,4

0400

In CO2 Calibration Mode (if CO2 sensor installed )2,4

0400

In SPAN Calibration Mode

0800

UNITS OF MEASURE FLAGS

 

UGM

0000

MGM

2000

PPB

4000

PPM

6000

SPARE/UNUSED BITS

1000, 8000

 

 

UNASSIGNED FLAGS (0000)

AZERO WARN2

 

DCPS WARNING

CANNOT DYN SPAN2

 

REAR BOARD NOT DET

CANNOT DYN ZERO3

 

SYNC WARNING1

CONC ALARM 13

 

SYSTEM RESET1

CONC ALARM 23

 

 

1

These status flags are standard for all instruments and should probably not be

2

modified.

 

 

Only applicable if the optional internal span gas generator is installed.

3

Only applicable if the analyzer is equipped with an alarm options.

3

It is possible to assign more than one flag to the same Hessen status bit. This

 

allows the grouping of similar flags, such as all temperature warnings, under the

 

same status bit.

 

 

Be careful not to assign conflicting flags to the same bit as each status bit will be triggered if any of the assigned flags is active.

182