Troubleshooting & Repair

Teledyne API – Technical Manual - Model 300E Family CO Analyzers

13.5.7.4.Status Outputs

The procedure below can be used to test the Status outputs:

1.Connect a jumper between the “D“ pin and the “” pin on the status output connector.

2.Connect a 1000 ohm resistor between the “+” pin and the pin for the status output that is being tested.

3.Connect a voltmeter between the “” pin and the pin of the output being tested (see table below).

Under the DIAGSIGNAL I/O menu (see Section 13.1.3), scroll through the inputs and outputs until you get to the output in question. Alternately turn on and off the output noting the voltage on the voltmeter, it should vary between 0 volts for ON and 5 volts for OFF.

Table 13-12: Status Outputs Check

PIN (LEFT TO RIGHT)

STATUS

1

SYSTEM OK

 

 

2

CONC VALID

3

HIGH RANGE

4

ZERO CAL

5

SPAN CAL

 

 

6

DIAG MODE

 

 

7

SPARE

 

 

8

SPARE

 

 

296

04288D DCN5752

Page 298
Image 298
Teledyne M300EM operation manual 12:Status Outputs Check