MODE AND

Calibration Procedures

Teledyne API – Technical Manual - Model 300E Family CO Analyzers

The following example sets sequence #2 to do a zero-span calibration every other day starting at 2:15 PM on September 4, 2008, lasting 15 minutes, without calibration. This will start ½ hour later each iteration.

Table 9-4: Example AutoCal Sequence

MODE AND

VALUE

COMMENT

ATTRIBUTE

 

 

SEQUENCE

2

Define Sequence #2

MODE

ZERO-SPAN

Select Zero and

Span Mode

TIMER ENABLE

ON

Enable the timer

STARTING DATE

Sept. 4, 2008

Start after

Sept 4, 2008

STARTING TIME

14:15

First Span starts at

2:15 PM

DELTA DAYS

2

Do Sequence #2

every other day

DELTA TIME

00:30

Do Sequence #2 ½

hr later each day

DURATION

30.0

Operate Span valve

for 15 min

CALIBRATE

ON

Calibrate at end of

Sequence

NOTE

The programmed STARTING_TIME must be a minimum of 5 minutes later than the real time clock for

setting real time clock (See Section 6.5.4).

Avoid setting two or more sequences at the same time of the day. Any new sequence that is initiated whether from a timer, the COM ports or the contact closure inputs will override any sequence that is in progress.

NOTE

With CALIBRATE turned ON, the state of the internal setup variables DYN_SPAN and DYN_ZERO is set

to ON and the instrument will reset the slope and offset values for the CO response each time the

AutoCal program runs.

This continuous readjustment of calibration parameters can often mask subtle fault conditions in the analyzer. It is recommended that, if CALIBRATE is enabled, the analyzer’s test functions, slope and offset values be checked frequently to assure high quality and accurate data from the instrument.

200

04288D DCN5752

Page 202
Image 202
Teledyne M300EM operation manual Mode And, Value, Comment, Zero-Span, Sept. 4, 30.0