Table 9-9 1660-Series Logic Analyzer Mainframe Module Event Status Register
Bit Bit Weig ht Bit Name C ondition
7 128 0 = not used
6 64 0 = not used
5 32 0 = not used
4 16 0 = not used
3 8 0 = not used
2 4 0 = not used
1 2 RNT 0 = Intermodule Run until not satisfied
1 = Intermodule Run until satisfied
0 1 MC 0 = Intermodule Measurement not satisfied
1 = Intermodule Measurement satisfied
Table 9-10 1660-Series Logic Analyzer Module Event Status Register
Bit Bit Weig ht Conditio n
7 128 0 = not used
6 64 0 = not used
5 32 0 = not used
4 16 0 = not used
3 8 1 = One or more pattern searches failed
0 = Pattern search es did not fail
2 4 1 = Trigger found
0 = Trigger not found
1 2 0 = Run until not satisfied
1 = Run until satisfied
0 1 0 = Measurement not satisfied
1 = Measurement satisfied
Mainframe Commands
MESR<N> (Module Event Status Register)
9–17