PROBe
Command :CHANnel<N>:PROBe <atten>
The PROBe command specifies the attenuation factor for an external probe
connected to a channel. The command changes the channel voltage
references such as range, offset, trigger level and automatic measurements.
The actual sensitivity is not changed at the channel input. The allowable
probe attenuation factor is an integer from 1 to 1000.
<N> An integer, from 1 to 2.
<atten> An integer from 1 to 1000
Example OUTPU T XXX;":CHAN1: PROB 10"
Query :CHANnel<N>:PROBe?
The PROBe query returns the probe attenuation factor for the selected
channel.
Returned Format [:CHANnel<N>:PROBe]<atten><NL>
Example OUTPUT XXX;":CH ANNEL1:PROBE ?"
CHANnel Subsystem
PROBe
29-7