OFFSet
Command :CHANnel<N>:OFFSet <value>
The OFFSet command sets the voltage that is represented at center screen
for the selected channel. The allowable offset voltage <value> is shown in
the table below. The table represents values for a Probe setting of 1:1. The
offset value is recompensated whenever the probe attenuation factor is
changed.
<N> An integer, from 1 to 2..
<value> allowable offset voltage value shown in the table below.
Vertical Range Offset Voltage
16 mV - 400 mV ±2 V
>400 mV - 1.6 V ±10 V
>1.6 V - 10 V ±50 V
>10 V - 40 V ±250 V
Example OUTPUT XXX;":CH AN1:OFFS 1.5"
Query :CHANnel<N>:OFFSet?
The OFFSet query returns the current value for the selected channel.
<N> An integer, from 1 to 2.
Returned Format [:CHANnel<N>:OFFSet] <value><NL>
Example OUTPUT XXX;":CH ANNEL1:OFFSE T?"
CHANnel Subsystem
OFFSet
29-6