THReshold
Command :MACHine{1|2}:SFORmat:THReshold<N>

{TTL|ECL|<value>}

The THReshold command allows you to set the voltage threshold for a given
pod to ECL, TTL, or a specific voltage from 6.00 V to +6.00 V in 0.05 volt
increments.
<N> {{1|2}|{3|4}|{5|6}|{7|8}} 1 through 8 for the HP 1660A, 1 through
6 for the HP 1661A, 1 through 4 for the HP 1662A, and 1 through 2 for the HP
1663A.
<value> Voltage (real number) 6.00 to +6.00
TTL Default value of +1.6 V
ECL Default value of 1.3 V
Example OUTPUT XXX;":MA CHINE1:SFORM AT:THRESHOLD 1 4.0"
Query :MACHine{1|2}:SFORmat:THReshold<N>?
The THReshold query returns the current threshold for a given pod.
Returned Format [:MACHine{1|2}:SFORmat:THReshold<N>] <value><NL>
Example OUTPUT XXX;":MA CHINE1:SFORM AT:THRESHOLD 4?"
SFORmat Subsystem
THReshold