SPERiod
Command :MACHine{1|2}:TWAVeform:SPERiod <sample_period>
The SPERiod command allows you to set the sample period of the timing
analyzer in the Conventional and Glitch modes. The sample period range
depends on the mode selected and is as follows:
2 ns to 8 ms for Conventional Half Channel 500 MHz
4 ns to 8 ms for Conventional Full Channel 250 MHz
8 ns to 8 ms for Glitch Half Channel 125 MHz
<sample_period> real number from 2 ns to 8 ms depending on mode
Example OUTPUT XXX;":MA CHINE1:TWAVE FORM:SPERIOD 50E9"
Query :MACHine{1|2}:TWAVeform:SPERiod?
The SPERiod query returns the current sample period.
Returned Format [:MACHine{1|2}:TWAVeform:SPERiod] <sample_period><NL>
Example OUTPUT XXX;":MA CHINE1:TWAVE FORM:SPERIOD ?"
TWAVeform Subsystem
SPERiod
23–18