RUNTil
Command :MARKer:RUNTil
{OFF|LT,<time>|GT,<time>|INRange,<time>,
<time>|OUTRange,<time>, <time>}
The RUNTil command allows you to set a stop condition based on the time
interval between the X marker and the O marker. In repetitive runs, when
the time specification is met, the oscilloscope stops acquiring data and the
advisory "Stop condition satisfied" will be displayed on screen.
<time> a real number specifying the time in seconds between the X and O markers
Example OUTPUT XXX;":MA RKER:RUNTIL LT ,1MS"
Query :MARKer:RUNTil?
The RUNTil query will return the current Run Until Time X - O (RUNTil)
setting.
Returned Format [:MARKer:RU NTil] {OFF|LT, <time>|GT,<t ime>|INRange ,<time>,
<time>|OUTRange,<time>,<time>}<NL>
Example OUTPUT XXX;":MA RKER:RUNTIL? "
MARKer Subsystem
RUNTil
31-11