Command :MARKer:SHO W {SAMPle| MARKer}
The SHOW command allows you to select either SAMPle rate or MARKer
data (when markers are enabled) to appear on the oscilloscope menus above
the waveform area.
The SAMPle rate or MARKer data appears on the channel, trigger, display,
and auto-measure menus. Marker data is always present on the marker
menu. While sample rate data is only present on the marker menu when time
markers are turned off.
Example OUTPUT XXX;":MA RKER:SHOW MARK ER"
TAVerage?
Query :MARKer:TAVerage?
The TAVerage query returns the average time between the X and O markers.
If there is no valid data, the query returns 9.9E37.
Returned Format [:MARKER:TAVERAGE] <time value><NL>
<time value> real number
Example OUTPUT XXX;":MA RKER:TAVERAG E?"
MARKer Subsystem