ACQMode
Command :MACHine{1|2}:TFORmat:ACQMode {TRANSitional

<size>|CONVentional <size>|GLITch}

The ACQMode (acquisition mode) command allows you to select the
acquisition mode for the timing analyzer. The options are:
conventional mode at full-channel 250 MHz
conventional mode at half-channel 500 Mhz
transitional mode at full-channel 125 MHz
transitional mode at half-channel 250 MHz
glitch mode.
<size> {FULL|HALF}
Example OUTPUT XXX;":MA CHINE2:TFORM AT:ACQMODE TRA NSITIONAL HALF "
Query :MACHine{1|2}:TFORmat:ACQMode?
The ACQMode query returns the current acquisition mode.
Returned Format [:MACHine{1|2}:TFORmat:ACQMode] {TRANSitional
<size>|CONVentional <size>|GLITch}<NL>
<size> {FULL|HALF}
Example OUTPUT XXX;":MA CHINE2:TFORM AT:ACQMODE?"
TFORmat Subsystem
ACQMode
21–5