Query :MARKer:OAUTo?
The OAUTo query returns the current settings.
Returned Format [:MARKer:OAUTo] CHANnel<N>,<type>
<level>,<slope>,<occurrence><NL>
Example OUTPUT XXX;":MA RKER:OAUTO?"
If <type> is not specified, the marker type will default to PERCent.
OTIMe
Command :MARKer:OTIMe <O marker time>
The OTIMe command moves the O marker to the specified time with respect
to the trigger marker.
<O marker
time>
time in seconds from trigger marker to O marker
Example OUTPUT XXX;":MA RKER:OTIME 1E- 6"
Query :MARKer:OTIMe?
The OTIMe query returns the time in seconds between the O marker and the
trigger marker.
Returned Format [:MARKer:OTIMe]<O marker time><NL>
Example OUTPUT XXX;":MA RKER:OTIME?"
MARKer Subsystem
OTIMe
31-10