The print query should NOT be sent with any other command or query on the
same command line. The print query never returns a header. Also, since
response data from a print query may be sent directly to a printer without
modification, the data is not returned in block mode.
Example
OUTPUT 707;":SYSTEM:PRINT? SCREEN"
SETupCommand :SYStem:SETup <block_data>
The :SYStem:SETup command configures the logic analyzer module as
defined by the block data sent by the controller. This chapter describes
briefly the syntax of the Setup command and query. Because of the
capabilites and importance of the Setup command and query, a complete
chapter is dedicated to it. The dedicated chapter is chapter 26, "DATA and
SETup Commands."
<block_data> <block_length_specifier><section>
<block_length_
specifier>
#8<length>
<length> The total length of all sections in byte format (must be represented with 8
digits)
<section> <section_header><section_data>
<section_
header>
16 bytes, described in the "Section Header Description" section in chapter 26.
<section_data> Format depends on the type of data
SYSTem Subsystem
SETup