VOTime?
Query :MARKer:VOTime? CHANNEL<N>
The VOTime query returns the current voltage level of the selected source at
the O marker.
Returned Format [:MARKer:VOTime]<level><NL>
<N> An integer from 1 to 2
<level> level in volts where the O marker crosses the waveform
Example OUTPUT XXX;":MA RKER:VOTIME? CHANNEL1"
For compatibility with older modules, the OVOLt query will function the same as
the VOTime query.
VRUNs?
Query :MARKer:VRUNs?
The VRUNs query returns the number of valid runs and the total number of
runs made. Valid runs are those where the edge search for both the X and O
markers was successful, resulting in valid marker time measurement.
Returned Format [:MARKer:VRUNs] <valid runs>,<total runs><NL>
<valid runs> positive integer
<total runs> positive integer
Example OUTPUT XXX;":MA RKER:VRUNS?"
MARKer Subsystem
VOTime?
31-16