Query :WLISt:OTIMe?
The OTIMe query returns the O Marker position in time. If data is not valid,
the query returns 9.9E37.
Returned Format [:WLISt:OTI Me] <time_valu e><NL>
<time_value> A real number
Example OUTPUT XXX;":WL IST:OTIME?"
RANGe
Command :MACHine{1|2}:WLISt:RANGe <time_value>
The RANGe command specifies the full-screen time in the timing waveform
menu. It is equivalent to ten times the seconds per division setting on the
display. The allowable values for RANGe are from 10 ns to 10 ks.
<time_value> A real number between 10 ns and 10 ks
Example OUTPUT XXX;":MA CHINE1:WLIST :RANGE 100E−9"
Query :MACHine{1|2}:WLISt:RANGe?
The RANGe query returns the current full-screen time.
Returned Format [:MACHine{1|2}:WLISt:RANGe] <time_value><NL>
<time_value> A real number between 10 ns and 10 ks
Example OUTPUT XXX;":MA CHINE1:WLIST :RANGE?"
WLISt Subsystem
RANGe
14–9