MINus
Command :DISPlay:MI Nus [<modu le_number> ,]<label>, <label>
The MINus command algebraically subtracts one channel from another and
inserts the resultant waveform to the display. The first parameter is an
optional module specifier. The module is identified by the slot number that
contains the oscilloscope card (always 2). The next two parameters are the
label of the waveform selected to be added to the display. The label names
are defined in the same manner as the INSert command.
<module_
number>
Always 2
<label> string of 1 alpha and 1 numeric character enclosed by single quotes
Example OUTPUT XXX;":DI SPLAY:MINUS 2, ’C1’,’C2’"
OVERlayCommand :DISPlay:OVERlay <label>,<label>
The OVERlay command overlays oscilloscope waveforms. The syntax
parameters are the labels of the waveforms that are to be overlaid. Only
waveforms sharing a common card can be overlaid. A label may be used only
once with each OVERlay command.
<label> string of 1 alpha and 1 numeric character enclosed by single quotes
Example OUTPUT XXX;":DI SPLAY:OVERLA Y ’C1’,’C2’"
DISPlay Subsystem
MINus
30-8