Fiber Testing

4

Calibrating the Tester

5On the Fiber Main Menu, press Calibration. The Fiber Calibration screen displays.

6Choose Double Ended, and press Next. The Fiber Detection screen displays.

7When the message below the diagram reads, “Ready to Calibrate,” press OK. The Calibration window opens.

8When the window reads, “Calibration Successful,” press OK in the window. The Fiber Main Menu displays. Remove the couplers and use the coupler ends of the jumpers when changing connections.

Leave the cable connections to the FiberProbes intact

Method B

Method B or the 1-jumper method is usually recommended for multimode fiber.