5Cable Testing in Expert Mode
The Expert Autotest Setup
recommended. Press Next to view the optional measurements you may want to add to your test. Add them by pressing the name or box.
3After you have made your optional choices, press Next to allow you to enable test result analysis functions. It is recommended that you should choose Manufacturer Specific Compensation when you are performing Category 6 channel testing. Press Next when done.
4The next Measurements Setup screen allows you to choose between Summary ELFEXT data and Complete ELFEXT data storage modes. If you have storage space and/or you will wish to recertify in the future using the stored data, the Complete ELFEXT Mode is recommended. When complete, press Next.
5The next Measurement Setup Screen allows the user to apply the 3 dB attenuation rule, which provides for special consideration of measurements where the attenuation is less than 3 dB or in certain ISO cases, 4 dB. This rule applies to both NEXT and Return Loss in the ISO case, but only to Return Loss in the TIA case. The FrameScope 350 automatically sets this choice based upon the standard chosen previously. It also implements the 4 dB
130 | Agilent FrameScope 350 User’s Manual |