5Cable Testing in Expert Mode

The Expert Autotest Setup

Setting the Maximum Frequency

The FrameScope 350 allows you to select the maximum range for swept frequency measurements. The choices depend somewhat on the standard selected previously, for example for Category 6, the 100 MHz choice is not available. To set the maximum frequency:

1In Expert Mode, press Autotest on the Main menu.

2On the Autotest Setup screen press Edit Settings. On the Autotest Settings screen press Max Frequency. On the Frequency Setup screen select the maximum frequency of the data being stored.

Setting the Plot Storage Requirements

Another storage choice is available, whether to save plots or not. Saving plots stores the entire swept frequency measurement. Summary data is the worst case value for each test and is the minimum required for standards compliance. To set the data being stored:

1In Expert Mode, press Autotest on the Main menu.

132

Agilent FrameScope 350 User’s Manual

Page 140
Image 140
Agilent Technologies N2610A user manual Setting the Maximum Frequency, Setting the Plot Storage Requirements