20. Tracking Oscillator Adjustment (Option 002)
Procedure
1.Remove the A10 Tracking Generator Assembly as described in Chapter 3. With the A10 Tracking Generator positioned next to the the HP 8560A, reconnect W14
2.Connect the equipment as shown in Figure
SPECTRUM ANALYZER | BNC TEE |
SJ14
Figure
3. Set the HP 8560A LINE switch on. Press (mCTRL), TRACKING GENRATOR , SRC PWR ON,
C E N T E R F R E Q . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ..300MHz SPAN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0 Hz
4. Press INSTR PRESET on the synthesized sweeper and set the controls as follows:
CW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.2107GHz P O W E R L E V E L . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . +12dBm FREQUENCY STANDARD Switch (rear panel) . . . . . . . . . . . . . . . . . . . . . . EXT
5. Set the frequency counter controls as follows:
SAMPLE RATE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Fully CCW 10
6.Remove the dust cap screw used to seal the tracking oscillator adjustment.
7.On the HP 856OA, press (m), TRACKING GEMRATOR , MORE 1 OF 3, MAN TRK ADJ , a, a. Rotate the knob counterclockwise until FINE TRACK ADJ is set to 0.