Cal Oscillator Unlock at Beginning of IF Adjust
1.Turn the spectrum analyzer luNE) switch off and then on. The words IF ADJUST STATUS appear on the display 10 seconds after the instrument is turned on (assuming the rest of the instrument is working correctly). Immediately observe the lower right corner of the display for error messages. If the message ERR 499 CAL UNLK appears (before errors ERR 561, 562 and 565), the cal oscillator is unable to
2.If the spectrum analyzer registers an unlocked cal oscillator, continue with step 3 to verify the presence of externally supplied signals.
3.Check A4U811 pin 9 for a 100 kHz
4.Check the $15 VF,
5.Check that A4U807 pin 5
Inadequate CAL OSC AMPTD Range
Refer to function block AC of A4 Calibration Oscillator Schematic Diagram in the HP 856OA/61B/63A Spectrum Analyzer Component Level Information binder.
1.If A4R826, CAL OSC AMPTD, has inadequate range to perform the IF Amplitude Adjustment, press LCAL).
2.Rotate A4R826 fully clockwise and disconnect W52 (coax 9) from A5J4.
3.Connect A5J4 to the input of a second spectrum analyzer.
4.Set the other spectrum analyzer controls as follows:
Center Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.7MHz Reference Level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5.Observe the spectrum analyzer display while pressing FULL IF ADJ . The signal level should be above
6.Rotate A4R826 fully counterclockwise. The signal should be below
7.Troubleshoot the ALC loop on this assembly using the following steps:
a.Connect a positive DVM probe to A4J9 pin 4.
b.On the spectrum analyzer press ICAL).
C. Press FULL IF ADJ . Observe the DVM reading between the displayed messages IF ADJUST STATUS: 300 kHz RBW and IF ADJUST STATUS: 3 kHz RBW. Duringthistime
period, the voltage should be within a 2 to 10 Vdc range.
d. Observe the DVM reading while IF ADJUST STATUS: AMPLITUDE is displayed. The reading should be within the 2 to 10 Vdc range.
IF Section