Table
HP 8560A | HP 8340A | Frequency of Offset PLL’s |
Zenter Frequency (MHz) CW Frequency (MHz) Sampling Oscillator (MHz)
389.5 | 4300.2 | 280.0 |
427.0 | 4337.7 | 282.5 |
449.5 | 4360.2 | 284.0 |
464.5 | 4375.2 | 285.0 |
479.5 | 4390.2 | 286.0 |
502.0 | 4412.7 | 287.5 |
509.5 | 4420.2 | 288.0 |
539.5 | 4450.2 | 290.0 |
569.5 | 4480.2 | 292.0 |
577.0 | 4487.7 | 292.5 |
599.5 | 4510.2 | 294.0 |
614.5 | 4525.2 | 295.0 |
629.5 | 4540.2 | 296.0 |
652.0 | 4562.7 | 297.5 |
659.5 | 4570.2 | 298.0 |
Some Synthesizer Section problems require placing the YTO PLL in an unlocked condition. This is done by moving jumper A14J23 to the TEST position. This grounds the YTO LOOP ERROR signal, disabling the CPU’s ability to detect an unlocked YTO. The FM Coil Driver’s output is set to its
Synthesizer Section troubleshooting is best done with the HP 8560A’s SPAN set to 0 Hz (even though it is still possible to sweep the Main and FM coils of the YTO).
With the YTO in its unlocked conditions and the SPAN set to 0 Hz, the nominal YTO frequency is not necessarily the value listed as LO FREQ in the Frequency Diagnose menu. The YTO has an initial pretune accuracy of 520 MHz, but this pretune accuracy shifts
f10 MHz as a function of the Sampler IF frequency. This enables the YTO loop to acquire lock when the Sampler IF is near its extremes. As a result, the nominal YTO frequency also shifts f10 MHz depending on the Sampler IF frequency.
The Roller Oscillator frequency is the same as the desired sampler IF. Table
nTo display the Roller Oscillator’s frequency press (CAL), MORE, FREQ DIAGNOSE ,
MAIN ROLLER. If the Sampler IF is negative (YTO frequency is lower than the desired Sampling Oscillator harmonic), the Main Roller frequency will be displayed as a negative number.
Synthesizer Section