When attached to the spectrum analyzer’s rear panel, the HP 85629B Test and Adjustment Module (TAM) p rovides diagnostic functions for the HP 8560A. Because the TAM connects directly to the analyzer’s internal data and address bus, it controls the analyzer’s hardware directly. It would be impossible to control the hardware to the same extent either from the analyzer’s front panel or over the
The TAM measures voltages at key points in the circuitry and flags a failure whenever the voltage falls outside the limits. The TAM locates the failure to a small functional area which can be examined manually.
Service KitThe HP 8560A Service Kit (HP part | number | service tools required to | ||||
repair | the instrument. Refer to Table | |||||
|
| Table |
| |||
|
|
|
|
|
|
|
|
| Description |
| Quantity |
| HP Part Number |
Cable | Puller |
| 1 |
| ||
PC Board | Prop |
| 1 |
| ||
Line | Filter | Assembly |
| 1 |
| |
Line | Switch Cable |
| 1 |
| ||
Extender Cable |
| 1 |
| |||
BNC to SMB |
| 2 |
| |||
|
|
|
|
|
|
|
Connector | Extractor Tool Kit |
| 1 |
|
Equipment required for operation verification, performance tests, adjustments, troubleshooting, and the Test and Adjustment Module is listed in Table
may | be substituted if it meets or exceeds the critical specifications listed | in the table. Refer to |
the | HP 856OA Installation and Verification Manual for the performance | tests. |