20. Tracking Oscillator Adjustment (Option 002)
8.Record the frequency counter reading in Table
9.On the HP 8560A, press MAN TRK ADJ , (255), (Hz. Rotate the knob clockwise until the FINE TRACK ADJ is set to 255.
10.Record the frequency counter reading in Table
11.Calculate Fcenter as shown below and record in Table
Fcenter= (Fl + F2) / 2
12.Set COARSE TRACK ADJ to 25. This sets the tracking oscillator near the center of its frequency range (the relationship between the COARSE TRACK ADJ dac number and the output frequency is nonlinear). Adjust COARSE TRACK ADJ and FINE TRACK ADJ until the frequency counter reads Fcenter flO0 Hz.
13.Record the values of COARSE TRACK ADJ and FINE TRACK ADJ in Table
14.Adjust AlOC3 TRK OSC CTR until the frequency counter reads 300 MHz f500 Hz.
15.Repeat steps 7 through 14 at least once more until no further adjustment of AlOC3 TRK OSC CTR is necessary.
16.Set the HP 8560A LINE switch off. Replace the dust cap screw on AlO. Disconnect all cables from AlO.
17.Reinstall A10 in the HP 8560A.
Table
TRACK ADJ
Adjustment Procedures