Table
Connector | Manual Probe Troubleshooting Test | Measured Signal Lines | ||
A2Jll | ADC/MUX Test | MSl, MS3 through MS6, MS8 | ||
| DAC Test | MS2, MS7, OS1 | ||
A2J201 | 10 Volt Reference Test | MS4 | ||
| Switch Drive Test | MS8 | ||
| Buffered X & Y DAC Outputs | MS2, MS7 | ||
| X Line Gen Test | MS6 | ||
| Y Line Gen Test | MS1 | ||
| Intensity | Offset Output | MS3 | |
A2J202 | Revision |
| MS1 | |
| X, Y, & Z Output Offset | MS3, MS4, MS7 | ||
| X | Output | Amplifier | MS3 |
| Y | Output | Amplifier | MS7 |
| Blanking | Test | MS8 | |
| Focus DAC Test | MS2 |
Blank Display
Use the following procedure if the instrument’s .display is blank. This procedure substitutes an
1.Connect the printer to the HP 8560A and set the printer’s address to the value required by the TAM. This is usually 1.
2.All of the
3.The
4.Connect the TAM’s probe cable to A2Jll.
5.Press
6.The yellow LED next to A2Jll should blink approximately ten times. If the LED fails to blink correctly, troubleshoot the digital section of the A2 Controller assembly.
7.Move the probe cable to A2J202. Press SOFT KEY #I and wait five seconds.
8.Press SOFT KEY #4 . The results should be sent to the printer.
9.Move the probe cable to A2J201, press SOFT KEY Xl and wait five seconds.