Procedure 16. A21 OCXO (Option 003)
Procedure 16. A21 OCXO (Option 003)Removal
1.Remove the
2.Place the analyzer on its
3.Fold out the Al4 and Al5 assemblies as described in Procedure 9, “Al4 and Al5 Assemblies Removal,” steps 3 and 4.
4.Remove the three screws (1) securing the OCXO to the main deck. See Figure
5.Disconnect W49, coax 82, and W50 from the OCXO and remove the OCXO from the analyzer.
Figure
Replacement
1.Connect W49, coax 82, and W50 to the OCXO and position the OCXO in the analyzer.
2.Secure the OCXO to the analyzer’s main deck using three screws (1). See Figure
3.Fold the Al4 and Al5 assemblies into the analyzer as described in Procedure 9.
4.Perform the
Assembly Replacement