4. IF Amplitude Adjustments
4. IF Amplitude AdjustmentsThe IF Amplitude Adjustments consist of the Cal Oscillator Amplitude adjustment and the Reference 15 dB Attenuator adjustment.
Assembly Adjusted
A4 Log Amp/Cal Oscillator
A5 IF Assembly
Related Performance Tests
IF Gain Uncertainty Test
Scale Fidelity Test
Equipment
Frequency Synthesizer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 3335A
Adapters
Type N (m) to BNC (f) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Cables
BNC,122cm (48in) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 10503A Test
T E S T C A B L E
/
I
IS P E C T R U M
IA N A L Y Z E R
F R E Q U E N C YI
I
Figure
Adjustment Procedures