14. External Mixer Bias Adjustment
14. External Mixer Bias AdjustmentAssembly Adjusted
Al5 RF Assembly
Related Performance Test
There is no related performance test for this adjustment procedure.
Description
A voltmeter is connected to the HP 8560A IF INPUT with the external mixer bias set to off. The bias is adjusted for a 0 Vdc output.
D I G I T A L V O L T M E T E R
ADAPTER
SPECTRUM
A l 4ANALYZER kRD
F R E Q U E N C Y /PROP ^ ^ - - ^
A l 5 R F |
| SK1 1 6 |
|
Figure
Equipment
DVM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 3456A
Adapters
Type BNC (f) to SMA (m) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Cables
BNC,122 cm(48in) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 10503A
Procedure
1.Set the HP 8560A (LINE) switch off and disconnect the ac power cord. Remove the analyzer cover and connect the equipment as illustrated in Figure
2.Set the HP 3456A controls as follows:
FUNCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . DCVOLTS R A N G E . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . O . l V RESOLUTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ..lOOmV
Adjustment Procedures